| Sign In | Join Free | My polstate.com |
|
| Categories | Melt Flow Indexer |
|---|---|
| Brand Name: | JHH |
| Model Number: | XNR-400C |
| Certification: | ISO9001, CE |
| Place of Origin: | Beijing, China |
| MOQ: | 1 set |
| Payment Terms: | T/T,L/C,D/A,D/P,Western Union,MoneyGram |
| Supply Ability: | 20 sets per month |
| Delivery Time: | 7-15 working days |
| Packaging Details: | Wooden case |
| Product Name: | High-Performance Melt Flow Index Tester for Plastic Materials with Dual Temperature Control and Integrated Printer |
| Temp.range: | 100-450℃ |
| Temperature Accuracy: | ±0.2°C |
| Temperature Resolution: | 0.1°C |
| Load: | All 21.6kg |
| Time range: | 0.1~999s |
| The diameter of the mold: | 2.095±0.005 mm |
| Cutting: | Manual, automatic |
| Company Info. |
| Beijing Tonghe Testing Machine Co., ltd |
| Verified Supplier |
| View Contact Details |
| Product List |
High-Performance Melt Flow Index Tester for Plastic Materials with Dual Temperature Control and Integrated Printer
Standards:
GB/T 3682-2000: Determination of Melt Flow Rate and MVR for
Thermoplastics
ISO 1133:1997: MFR/MVR Testing of Thermoplastic Melts
ASTM D1238: Extrusion Plastometer Method for Thermoplastic Melt
Flow Rate
Product Features:
Technical Parameters:
| Temperature range | RT ~ 450 ℃ |
| Temperature fluctuation | ± 0.2 ℃ |
| Temperature uniformity | ± 1 ℃ |
| Temperature display resolution | 0.1 ℃ |
| Time display resolution | 0.1S |
| Barrel diameter | Φ2.095 ± 0.005mm, Φ1.05mm±0.005mm |
| Outlet length | 8.000 ± 0.025mm |
| Loading barrel diameter | Φ9.550 ± 0.025mm |
| Weight accuracy | ± 0.5% |
| Displacement measurement accuracy | 0.01mm |
| MVR timing accuracy | 0.01S |
| Displacement measurement encoder | Omron photoelectric rotary encoder |
| Test range of MFR | 0.1-150g / 10min |
| Test range of MVR | 50-5000cm3 / 10min |
| Cutting method | manual and automatic cutting in one machine |
| Test load | 8 levels in total, 8 sets of weights |
| Power supply voltage | AC220V ± 10% 50 / 60HZ, 1ph |



|